2201R Mobile Tester SCPI Reference Guide
Chapter 5 System commands
47090/323
Copyright © 2016 Aeroflex Limited. All rights reserved.
52
Introduction
System parameters, such as the number of unread error messages, the test set's interface parameters etc. may be read
out or set using the commands of this subsystem.
System subsystem
:SYSTem:ERRor[:NEXT]?
Syntax
:SYSTem:ERRor[:NEXT]?
Parameters
There are no parameters.
Description
There is solely a query form of this command available.
Query
Returns the oldest unread error message from the internal error queue of the 2201R. The queue entry returned is a
string containing the error no. and additional text. The maximum length of the string is 255 characters.
Note: An overview of all SCPI error messages can be found in section SCPI command errors.
Example
:SYSTem:ERRor:NEXT?
String returned: 113,"Undefined header;(0)^SCPIParseCommand: SYSTEM:WRONG:CMD "
:SYSTem:ERRor:COUNt?
Syntax
:SYSTem:ERRor:COUNt?
Parameters
There are no parameters.
Description
There is solely a query form of this command available.
Query
Returns the number of unread error messages in the internal error queue of the 2201R. The string returned contains
one integer. The maximum number of errors stored internally is 10.
Example
:SYSTem:ERRor:COUNt?
String returned: "0"
This means that there are no unread error messages in the error queue.