2201R Mobile Tester SCPI Reference Guide
Chapter 6 Status commands
47090/323
Copyright © 2016 Aeroflex Limited. All rights reserved.
65
:STATus:OPERation:ENABLe
Syntax
:STATus:OPERation:ENABLe <int1>
Parameters
int1 is an integer. The minimum value for int1 is 0, the maximum value is 32767. The default value is 0.
Description
This command sets the Enable filter of the General operation register group. This mask is ANDed with the general
operation event register and thus decides what kind of events are forwarded to bit 7 of the service register.
Note: The service register is often also referred to as the status byte register or status byte.
Query
There is no query form of this command available.
Example
:STATus:OPERation:ENABle 129
This means that any setting of bits 1 (calibrating) or 7 (correcting) of the general operation event register raises bit
7 of the service register.
:STATus:OPERation:PTRansition
Syntax
:STATus:OPERation:PTRansition <int1>
Parameters
int1 is an integer. The minimum value for int1 is 0, the maximum value is 32767. The default value is 32767.
Description
Sets the PTR mask of the General operation register group. This mask is ANDed with the up to 15 bits of the PTR
transition filter of the general operation condition register to allow a positive transition of any bit (i.e. a transition
from logic 0'to'1')toreachthegeneraloperationeventregister.'
Notes
- The default of this mask is 32767 — that means that the mask allows any positive transition of the lower 15 bits of
the condition-type register to reach the event-type register.
- Bit 16 of the general operation condition register is not used.
Query
There is no query form of this command available.
Example
:STATus:OPERation:PTRansition 0
This means that none of the positive transitions of the condition-type register are forwarded to the event-type
register.