91e_Trouble shooting en bedingt.fm, 10.09.2010
P1890E/EN 2010-08
47
Troubleshooting
9
RF communication is
partly interrupted.
Distance between base
station and tool is too
great.
If channel 1 is selected, the distance can be up to
30 m. If channel 2 or 3 is selected, the distance can
be up to 10 m.
➔
Move the tool close to the base station to check
whether communication is successful. If so,
increase output power on base station and on
the tool, or move the base station closer to the
tool.
Output power is too low.
➔
Increase the output power of the base station
and of the tool. If channel 1 is selected, you can
choose up to 25 mW for the output power. If
channel 2 or 3 is selected, you can choose 1
mW for the output power of the base station and
5 mW for the output power of the tool.
Too much traffic on the
same channel.
➔
Reduce traffic on the wireless network. Deacti-
vate torque plot data transmission.
Too many tools on the
same channel.
➔
Use different channels for different base sta-
tions.
Other 2,4 GHz devices
on the same frequency.
➔
Use a different channel.
Distance for RF commu-
nication is too short
Output power is too low.
➔
Increase the output power of both the base sta-
tion and the tool.
Location of the base sta-
tion bad.
➔
Move the base station to a location where there
is an unobstructed line of view between the
base station and the tool.
Problem
Possible cause
Action
Barcode scanner on tool
Barcode scanner does
not activate when the
start switch is pressed.
Parameter for Part-ID is
not set to Enable Inter-
locked.
➔
In the
Communication/Part-ID
screen of the
controller, check that the Enable parameter is
set to Enable Interlocked.
Barcode has already
been read.
➔
Activate a further read cycle in the scanner
submenu.
➔
Press the left function key on the tool in order
to activate another read cycle.
Note:
Only available if the parameter for <F1> but-
ton on Tool is set to Read barcode in the
Advanced Application Builder / System Settings
screen of the controller.
Barcode is not read.
Barcode scanner win-
dow is not clean.
➔
Clean the window by using a damp cloth and a
conventional window cleaner.
Barcode type is dis-
abled by parameter set-
ting.
➔
In the
Communication/Part-ID
screen, check
that parameter Barcode Type is set to the
appropriate barcode type.
Problem
Possible cause
Action
RF15.4 data communication between controller and tool
Содержание 17BP B
Страница 4: ...4 P1890E EN 2010 08 90a_ Deckblatt en fm 10 09 2010...
Страница 20: ...20 P1890E EN 2010 08 91b_1_5 en bedingt fm 10 09 2010 First Operation 6...
Страница 48: ...48 P1890E EN 2010 08 91e_Trouble shooting en bedingt fm 10 09 2010 Troubleshooting 9...
Страница 54: ...54 P1890E EN 2010 08 90e_Ersatzteile en fm 10 09 2010 Spare parts 10...