CTD-N Technical Manual
P/N 95F-6000-00 (February 2017)
page 9
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
Conductivity
Sensor type:
Non-external inductive cell
Operational Range:
0 to 70 mS/cm
Typical Accuracy:
1,2
±0.003 mS/cm @ 35 PSU, 22°C
Typical Drift:
1,2,3
±0.001 mS/cm/month
Thermal Stability
4
:
±.003 mS/cm/°C; Δ°C rel. to 22 °C (typical)
Resolution:
0.0001 mS/cm
1
Specified at 22
°
C and 35PSU - Defined as the root sum of the squares (RSS) of endpoint non-linearity,
repeatability error and calibration uncertainty.
2
Specified from 0-70 mS/cm
3
Over 1 year (Typical)
4
ΔT rel. to 22°C (Typical)
Temperature
Sensor type:
Aged thermistor
Range:
-5 to 35 °C
Accuracy:
±0.005°C
Stability:
±0.0005°C/month
Resolution:
0.0001°C
Response:
100 msec
Pressure
Sensor type:
Precision micro-machined silicon transducer
Range:
Customer specified
Accuracy:
±0.05% of full scale
Stability:
±0.004% of full scale/month
Resolution:
0.001% of full scale
Response:
25 msec