CTD-N Technical Manual
page 112
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
***I
Go to Interval mode
Usage
***I <filename> <CR> or <LF>
Operating Modes
Open
Description
The ***I command creates a file in the instrument and sets the instrument into the Interval operat-
ing mode of the Run mode if valid interval and on times have been entered. In Interval mode the
instrument turns on and acquires and saves data to the file at the beginning of each interval time as
set by the ITIME command for the on time as set by the OTIME command. Data are recorded at the
sample rate set by the SRATE command. The instrument turns off at the end of each on time.
Returns
<CR><LF> Time = 10:51:41.54 Date = 08-12-2005 This ON Time Ends 11:11:40 08-12-2005 The
NEXT Interval Starts 11:51:40 08-12-2005
Entering a <CR> or <LF> character a second time and every time thereafter displays a scan of data
during the on time only followed by a <CR><LF>.
See Also
ITIME, OTIME, S, SC, SRATE