Go to the Run operating mode
Usage
***R <CR> or <LF>
Operating Modes
Open
Description
The ***R command sets the instrument into the Run operating mode. In Run mode the instrument
acquires data in Continuous, Interval or Delayed Start mode. Data are recorded at the sample rate
set by the SRATE command.
In Continuous mode, data are acquired continuously.
For the instrument to acquire data in Interval mode, valid interval and on times must have been en-
tered. In Interval mode the instrument turns on and data are acquired at the beginning of each in-
terval time as set by the ITIME command for the on time as set by the OTIME command. The in-
strument turns off at the end of each on time.
For the instrument to acquire data in Delayed Start mode, a valid delayed start time and date must
have been entered. In Delayed Start mode data are acquired in either Continuous or Interval mode
commencing at the time specified by the DTIME command and the date specified by the DDATE
command.
Returns
<CR><LF>
Entering a <CR> or <LF> character a second time and every time thereafter displays a scan of data
followed by a <CR><LF>.
See Also
ITIME, OTIME, DTIME, DDATE, S, SC, SRATE