CTD-N Technical Manual
page 44
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
16. Click OK.
The Delayed Start Continuous window and the CTD-N Configuration dialog box close. Data
will continue to be acquired and stored until data logging is stopped, the memory is full, or
the battery pack is depleted.
17. Disconnect the test cable from the instrument and install the dummy plug.
CAUTION.
Failure to install the dummy plug will cause the connector pins to corrode and
may cause damage to the instrument.
18. Deploy the instrument in the required manner.
Configuring and Deploying the CTD-N in Delayed Start Interval Mode
To configure a CTD-N for deployment and to acquire and store data in Delayed Start Interval mode:
1. If necessary, replace the battery pack in the CTD-N. (See
for instructions on how to replace the battery pack.)
NOTE.
Replacing the battery pack will not affect the instrument’s configuration, stored data,
or time and date settings.
2. Connect the CTD-N to your computer.
3. Choose Start
➤
All Programs
➤
RD Instruments
➤
CTDPro.
CTDPro starts and the CTDPro Main window opens.
4. Choose CTD
➤
Configure and Deploy.
If the CTD-N is currently logging data, a window opens asking if you want to stop logging:
5. Click Yes to stop data logging, or click No to continue.