CTD-N Technical Manual
page 124
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
CCOP, RCOP, SCOP
Sets, clears, and reads continuous on power up
Usage
CCOP
RCOP
SCOP
Operating Modes
Open
Description
These commands set, clear, and read continuous data output after power up. The SCOP command
enables continuous output of data immediately after power up. Data are output at the instrument
sampling rate as set by the SRATE command and the baud rate as set by the SB command. The
CCOP command clears the continuous data on power up. RCOP reads the state of continuous data.
NOTE.
A “***E” command must be issued after either command for changes to take effect
upon next power up.
Returns
The state of continuous data output on power up.
Example
1. SCOP<CR><LF>
Continuous set<CR><LF>
***E<CR><LF>
2. CCOP<CR><LF>
Continuous cleared<CR><LF>
***E<CR><LF>
3. RCOP<CR><LF>
Continuous set<CR><LF>
-OR-
Continuous cleared<CR><LF>
See Also
***E, SRATE, SB