14
Chapter 1 Use and maintenance: reminders and further information
Maintenance manual
2016/02
AlveoPC
FREQUENCY
■
All the control test results (curves and computed values) must be
precisely kept.
We recommend keeping a hard copy in addition to a backup on the
PC.
■
The following frequency for these control tests is recommended:
❏
Every 100 tests or after 1 month (whichever is the earlier).
Note:
Users of Alveo instruments are strongly recommended to
be members of a laboratory comparison round robin such
as BIPEA, AACC,…; in which case the rules of this circuit
should be observed.
❏
After a major overhaul of instruments, for instance after:
- replacing the mixing blade,
- adjusting the extraction passage,
- realigning the mixer's bowl and gear motor,
- replacing the Alveo spacers,
- cleaning the Alveo pneumatic measuring circuit,
- replacing an Alveograph pressure sensor.
❏
If in doubt about any measurements.
CONTROL SAMPLE
■
Quantity
An adequate quantity of sample flour must be kept for a sufficient
period to monitor the performance of the instruments.
■
Storage conditions
❏
Initial characterization
The characteristics of the sample must be measured carefully. In
particular, accurately measure:
❏
the moisture content of the sample,
❏
its alveographic characteristics (a minimum of 3 tests),
Reminder:The milled sample must be stored for at least 6 days after
production, and for no more than 15 days, at room tem-
perature (20°C) in a well-ventilated environment.
■
Storage
The control sample must be kept:
❏
In a clean, dry container made of inert material, in a refrigerated
room at constant temperature, and must be labelled,
❏
for a period not exceeding admissible limits
1
.
Note:
It can be stored at low temperature
(–20°C for instance); to that end, the usual freezing rules
must be observed.
1.
The maximum storage time is 2 months at 4°C.
Содержание AlveoPC
Страница 2: ......
Страница 4: ...AlveoPC 4 Maintenance manual...
Страница 6: ...AlveoPC 6 Maintenance manual...
Страница 7: ...7 Maintenance manual 2016 02 AlveoPC Chapter 1 USE AND MAINTENANCE REMINDERS AND FURTHER INFORMATION...
Страница 8: ...8 Maintenance manual 2016 02 AlveoPC...
Страница 16: ...16 Chapter 1 Use and maintenance reminders and further information Maintenance manual 2016 02 AlveoPC...
Страница 17: ...17 Maintenance manual 2016 02 AlveoPC Chapter 2 TROUBLE SHOOTING...
Страница 18: ...18 Maintenance manual 2016 02 AlveoPC...
Страница 58: ...58 Chapter 2 Trouble shooting Maintenance manual 2016 02 AlveoPC...
Страница 59: ...59 Maintenance manual 2016 02 AlveoPC Chapter 3 REPAIRS...
Страница 60: ...60 Maintenance manual 2016 02 AlveoPC...
Страница 66: ...66 Chapter 3 Repairs Maintenance manual 2016 02 AlveoPC 2 Instructions See following pages...
Страница 67: ...Chapter 3 Repairs 67 Maintenance manual 2016 02 AlveoPC A Alveograph...
Страница 111: ...111 Maintenance manual 2016 02 AlveoPC B Mixer...
Страница 112: ...112 Maintenance manual 2016 02 AlveoPC...
Страница 146: ...146 Maintenance manual 2016 02 AlveoPC...
Страница 147: ...147 Maintenance manual 2016 02 AlveoPC Chapter 4 SPARE PARTS...
Страница 148: ...148 Maintenance manual 2016 02 AlveoPC...
Страница 150: ...150 Chapter 4 Spare parts Maintenance manual 2016 02 AlveoPC...
Страница 152: ...152 Chapter 4 Spare parts Maintenance manual 2016 02 AlveoPC figure 1 Alveograph 1 2 5 3 4...
Страница 154: ...154 Chapter 4 Spare parts Maintenance manual 2016 02 AlveoPC figure 2 Alveograph assembly 5 1 2 3 4...
Страница 162: ...162 Chapter 4 Spare parts Maintenance manual 2016 02 AlveoPC figure 6 Alveograph Lower plate assembly 5 2 6 4 3 7 1...
Страница 174: ...174 Chapter 4 Spare parts Maintenance manual 2016 02 AlveoPC...