INST.No.INE-378-P9CE
- 40 -
6.9 Overflow/underflow
● “
oFL
” is displayed when the measured temperature is higher than the maximum value of measuring
range + 20ºC.
● “
uFL
” is displayed when the measured temperature is lower than the minimum value of measuring
range – 20ºC.
6.10 Under flow display for two-color type
radiation thermometer only:
“CLP”
“CLP” is displayed when the measured temperature using two-color type radiation thermometer is lower
than minimum value of measuring range. And analog output is lower than 4mA.
This phenomenon occurs to etc. (1) View lacking (2) Measuring with cap to tip of fiber optics
Reference
・
When measured temperature has not reached to the minimum value of measuring
range at the post period of measurement start, CLP' or 'UFL' are displayed.
・
The view lacking is expected, in the case that this phenomenon occurred although
the temperature of the measurement objects is rising.
[ Overflow ]
[ Underflow ]