INST.No.INE-378-P9CE
- 36 -
6.5.10 Communications programming
(
for IR-FA
S
with optional communications
)
This mode is for the programming of the address (effective in multi-drop connections) of this
thermometer and communications speed.
6.5.10-1 Address programming
1) To program the address of this thermometer, press
SEL
key in
the engineering mode screen until “
Adr
” appears in the sub
display as shown in the right figure.
2) By pressing
key, the least significant digit of the main
display will blink.
3) Use
△
or
▽
key to program the desired value.
4) To activate this programming, press
ENT
key.
6.5.10-2 Communication speed selection
1) To select the communications speed, press
SEL
key in the
engineering mode screen until “
SPd
” appears in the sub display
as shown in the right figure.
2) By pressing
key, “
4.8
”, “
9.6
” or “
19.2
” will blink in the
main display.
3) Use
△
or
▽
key to select the desired type.
4) To activate this selection, press
ENT
key.
[Communications speed]
4.8
: 4800 bps
9.6
: 9600 bps
19.2
: 19200 bps
• The numeric figure in the main display is available from 1 to 99. However, the
programmable range is
1
to
32
.
• The default value is
1
.
R e m a r k s
• The numeric figure in the main display is available from
1
to
99
.
However, the programmable range is
1
to
32
.
• The default value is
1
.
R e m a r k s
Tb
CONT
MEM
PEAK
AL
AH
C
F
ε
[ Address ]
Sub display
Main display
Sub display
Sub display
Tb
CONT
MEM
PEAK
AL
AH
C
F
ε
[ Communications speed ]
Main display
Caution
To return to the measurement mode screen from the engineering mode screen
• The screen will be automatically returned to the measurement mode screen if any
key is not pressed for 1 second in the programming of each parameter.
• The screen will return to the measurement mode screen when the programming from
[6.5.1 Analog output scaling] to [6.5.8 Analog output correction] is completed.
Note
: Up to [6.5.9 Analog input programming] in case of IR-FA