INST.No.INE-378-P9CE
- 32 -
6.5.8-5 Selection of analog output correction for programming item
This mode is for the selection of continuous entering of analog output correction data, execution of
correction command or exiting from this programming.
1) After the entering of the data after the analog output correction
in [6.5.8-4 Entering of data after analog output correction],
press
SEL
key until “
End
” (selection screen of analog output
correction programming item) appears in the sub display as
shown in the right figure.
2) By pressing
key, the most significant digit of the main
display will blink.
3) Use
△
or
▽
key to program the desired value.
4) To shift to the next lower digit, press
key.
The next lower digit will blink.
5) Repeat the above procedure of (3) and (4) up to the least
significant digit for programming (4 digits).
6) To activate this programming, press
ENT
key.
[Correction programming items]
Cont :
For continuous entering. Press
SEL
key to return to the
Programming screen of [6.5.8-2 Programming of analog
output correction data No.]
CAnc :
For exiting from this programming. Press
SEL
key to
move to the next mode.
Go
:
For execution of correction command.
• The default is
cont
(continuous entering).
R em a rk s
Main display
Tb
CONT
MEM
PEAK
AL
AH
C
F
ε
[Continuous data entering]
Sub display
[Exit from programming]
Tb
CONT
MEM
PEAK
AL
AH
C
F
ε
Sub display
Main display
Tb
CONT
MEM
PEAK
AL
AH
C
F
ε
[Correction command execution]
Main display
Sub display