INST.No.INE-378-P9CE
- 21 -
6.5 Engineering mode
In the measurement mode, press
SEL
and
ENT
keys simultaneously to display the engineering mode screen.
Refer to the engineering mode screen in [6.3 Parameter programming/selection (Engineering mode)].
6.5.1 Analog output scaling programming
6.5.1-1 Analog output minimum value programming…"OutL"
This programming is for the minimum value of the temperature scaling when the measured value is
outputted by analog signal.
1) To program the analog output minimum value, press
SEL
key
in the engineering mode screen until “
OutL
” appears in the
sub display as shown in the right figure.
2) By pressing
key, the most significant digit of the main
display will blink.
3) Use
△
or
▽
key to program the desired value.
4) To shift to the next lower digit, press
key.
The next lower digit will blink.
5) Repeat the above procedure of (3) and (4) up to the least
significant digit for programming (4 digits).
6) To activate this programming, press
ENT
key.
6.5.1-2 Analog output minimum value programming…"OutH"
This programming is for the minimum value of the temperature scaling when the measured value is
outputted by analog signal.
1) To program the analog output maximum value, press
SEL
key
in the engineering mode screen until “
OutH
” appears in the
sub display as shown in the right figure.
2) By pressing
key, the most significant digit of the main
display will blink.
3) Use
△
or
▽
key to program the desired value.
4) To shift to the next lower digit, press
key. The next lower
digit will blink.
5) Repeat the above procedure of (3) and (4) up to the least
significant digit for programming (4 digits).
6) To activate this programming, press
ENT
key.
• The programmable range is
0
to
6280ºC
.
• The defaults is
OutL
(Minimum value) : Minimum value of the measuring range
Reference
• The programmable range is
0
to
6280ºC
.
• The defaults is:
OutH
(Maximum value) : Maximum value of the measuring range.
Reference
Tb
CONT
MEM
PEAK
AL
AH
C
F
ε
[Maximum value]
Main display
Sub display
Caution
To return to the measurement mode screen from the engineering mode screen
• The screen will be automatically returned to the measurement mode screen if any
key is not pressed for 1 second in the programming of each parameter.
• The screen will return to the measurement mode screen when the programming from
[6.5.1 Analog output scaling programming] to [6.5.8 Analog output correction
programming] is completed.
Note: Up to [6.5.9 Analog input programming] in case of IR-FA