- 20 -
6.4.5 Alarm programming
This programming is for the selection of the alarm type (high alarm, low alarm or alarm disable) and the
programming an alarm set-point (1 point only in the high alarm or the low alarm).
6.4.5-1 Alarm type selection (High alarm, low alarm or alarm disable)
1) To select the alarm type, press
SEL
key in the measurement
mode screen until “
A.Mod
” appears in the sub display as
shown in the right figure.
2) By pressing
key, “
no
”, “
Hi
” or “
Lo
” will blink in the main
display.
3) Use
△
or
▽
key to select the desired type.
4) To activate this selection, press
ENT
key.
[
Alarm type]
No: Alarm disable
Hi: High alarm (The open collector of the contact output is ON when the measured value becomes
over the alarm set-point)
Lo: Low temperature alarm (The open collector of the contact output is OFF when the measured
value becomes below the alarm set-point)
6.4.5-2 Alarm set-point (1 point only in high alarm or low alarm in
[6.4.5-1 Alarm type selection]
1) To program the alarm set-point, press
SEL
key in the
measurement mode screen until “
ALM
” appears in the sub
display as shown in the right figure.
2) By pressing
key, the most significant digit of the main
display will blink.
3) Use
△
or
▽
key to program the desired value.
4) To shift to the next lower digit, press
key.
The next lower digit will blink.
5) Repeat the above procedure of (3) and (4) up to the least
significant digit for programming (4 digits).
6) To activate this programming, press
ENT
key.
• The programmable range is
0
to
6280ºC
.
• The defaults are:
Hi
(High alarm) : Maximum value of the measuring range
Lo
(Low alarm) : Minimum value of the measuring range
R em ar k s
Tb
CONT
MEM
PEAK
AL
AH
C
F
ε
[Low alarm]
Tb
CONT
MEM
PEAK
AL
AH
C
F
ε
[High alarm]
Sub display
Main display
Sub display
Main display
Sub display
Tb
CONT
MEM
PEAK
AL
AH
C
F
ε
Main display
Tb
CONT
MEM
PEAK
AL
AH
C
F
ε
[Alarm disable]
Sub display
Main display