INST.No.INE-378-P9CE
- 34 -
6.5.9-2 Emissivity scaling for remote emissivity programming
This mode is for the scaling of the emissivity value when the remote emissivity programming is selected.
1) To program the minimum emissivity value for the scaling, press
SEL
key in the engineering mode screen until “
AinL
” appears
in the sub display as shown in the right figure.
2) By pressing
key, the most significant digit of the main
display will blink.
3) Use
△
or
▽
key to program the desired value.
4) To shift to the next lower digit, press
key.
The next lower digit will blink.
5) Repeat the above procedure of 3) and 4) up to the least
significant digit for programming (4 digits).
6) To activate this programming, press
ENT
key.
7) To program the maximum emissivity value for the scaling,
press
SEL
key in the engineering mode screen until “
AinH
”
appears in the sub display as shown in the right figure.
8) Repeat the above procedure from 2) to 6) .
• The programmable range is
1.999
to
0.001
.
• The defaults are:
AinL
(Minimum value) :
0.050
AinH
(Maximum value) :
1.999
Remarks
[ Minimum value of
emissivity scaling]
Tb
CONT
MEM
PEAK
AL
AH
C
F
ε
Sub display
Main display
[ Maximum value of
emissivity scaling]
AL
AH
C
F
ε
Tb
CONT
MEM
PEAK
Sub display
Main display
Caution
Confirmation of the value programmed by the remote emissivity programming.
In the emissivity programming screen in the operator mode, you can confirm the value
programmed by the remote emissivity programming. However, even if you change the
analog input value on the condition that the programmed value is displayed, the
displayed value is not changed. For your confirmation of the new value, it is necessary
to return to the measurement mode and then display the emissivity programming
screen again.