INST.No.INE-378-P9CE
- 24 -
6.5.3-2 Peak hold reset type selection
(It is necessary to select when the peak hold is selected in [6.5.3-1 Hold function type selection].)
1) After the selection of “
PEAk
” in [6.5.3-1 Hold function type
selection], press
SEL
key until “
HrSt
” (selection screen of
peak hold reset type) appears in the sub display as shown in the
right figure.
2) By pressing
key, “
no
”, “
in
” or “
Et
” will blink in the main
display.
3) Use
△
or
▽
key to select the desired type.
4) To activate this selection, press
ENT
key.
[Reset types]
no
: Reset disable
in
: Internal (time reset)
Et
: External (remote
contact reset)
For the sample hold, the measured value
is hold by the remote contact.
(The pulse width more than 55msec is
required for the remote contact.)
C au t io n
• The default is no (reset disable).
Remarks
C au t io n
T
R
: Reset time 0.0 to 99.9 sec
In case of “0.0” is set, hold time is less
than 0.0 to 0.1.
Tb
CONT
MEM
PEAK
AL
AH
C
F
ε
[Reset disable]
Main display
Sub display
Tb
CONT
MEM
PEAK
AL
AH
C
F
ε
[External reset]
Main display
Sub display
Tb
CONT
MEM
PEAK
AL
AH
C
F
ε
[Internal reset]
Main display
Sub display
【
Peak hold (Reset disable)
】
Input
T
R
T
R
【
Peak hold internal (Time reset)
】
【
Peak hold external (Remote contact reset)
】
Output
Input
Output
Input
Output
External
contact output