43
Mobile 2 RPW User Manual
8.10 TROUBLE -SHOOTING
Fault description
Possible cause
Corrective actions
No power output
Leaks in handpiece cable or cable not
properly connected
Blocked or worn projectile
Handpiece defective
No projectile
Guide tube installed the wrong way round
Check the cable and tube connections
and replace them, if necessary
Dismantle the handpiece
Overhaul the handpiece
Replace the handpiece
Install a projectile
Turn the guide tube around
Irregular frequency
2 projectiles
Remove one projectile
Leak in the handpiece connector
Defective or missing red O-ring on plug
Send in handpiece or inform Service
8.11 ACCESSORIES AND SPARE PARTS
Handpiece overhaul kit
23825
Shaft cushion
22376
Handle cushion
22375
Shock transmitters
Detailed ordering information for the complete shock transmitter range is available from your local dealer.