7. Technical Specifications of the Control Device
7.1 DEVICE
R-SW operating mode
Single shock, continuous shock
1.0 - 4.0 bar / max. 21 Hz
4.6 bar / max. 15 Hz
5.0 bar / max. 12 Hz
V-ACTOR operating mode
1.4 - 3.0 bar / max. 31 Hz
3.8 bar / max. 22 Hz
5.0 bar / max. 13 Hz
Mains input voltage
100 - 240 VAC
Mains frequency
50 - 60 Hz
Mains fuse
T4AL/250 VAC
Power consumption
max. 200 VA
Compressed air output
1 - 5 bar
Ambient temperature during operation
10 - 40 °C
Ambient temperature during storage and transport
0° – 60° C frost-free
Ambient air pressure during operation
800 - 1060 hPa
Ambient air pressure during storage and transport
500 - 1060 hPa
Air humidity
5 - 95%, non-condensing
Control device weight
8.2 kg
Housing dimensions (W x H x D)
490 x 290 x 400 mm
Classification according to MDD
Class IIa device
Protection against the ingress of water
IPX1
Subject to technical modifications
Equipment safety (“essential performance“) according to IEC 60601-1, 3rd edition:
ME equipment shall be free from excessive pressure wave energy.
NOTE
When the medical product is distributed to third parties, the following must be observed:
– The complete device documentation must be delivered together with the medical product.
– The medical product may only be exported to a foreign country when the medical product and the corresponding indications
are allowed there.
7.2 TYPE PLATE
Fig. 7 - 1 Type plate