3. Operation
3.5 TREATMENT
CAUTION
The transport bag is provided only to transport the device. If the device is left in the transport bag during
treatment, the device becomes hot, due to lack of ventilation. Burns, conflagration and damages of the
device are possible.
•
Take the device out of the transport bag during treatment.
CAUTION
Read chapter 3.1 GENERAL WARNINGS AND SAFETY INFORMATION before beginning treatment.
Please also follow the instructions in chapter 8 and 9 for your handpiece.
Each time after the instrument has been transported, make sure that all functional checks have been
performed on the instrument before you start treatment.
Only perform treatments approved by the manufacturer!
To avoid safety hazards, use of the instrument for applications other than those specified in chapter 1.1.1
INDICATIONS is not allowed!
All status and error messages signaled during treatment must always be attended to
without delay!
The maximum energy level used during treatment must not cause the patient undue pain under any
circumstances.
CAUTION
We recommend that the user and the patient wear suitable hearing protection.
•
Always offer the patient hearing protection.
•
Apply a sufficient amount of coupling gel to the patient’s skin in the treatment area and to the shock
transmitter.
•
Do not apply more than 300 shocks to the same spot during treatment.
•
Avoid excessive pressure of the shock transmitter to the patient’s skin. Excessive pressure is not necessary
for the success of the treatment.
CAUTION
The shock transmitter surface temperature can reach up to 47°C. Extended skin contact can lead to minor
burns!
•
Interrupt treatment after a maximum of 6,000 pulses.
CAUTION
The handpiece may not be operated while idling (without an impact surface).
•
Do not trigger pulses unless the pulse transmitter is in contact with the treatment zone!