50
Mobile 2 RPW User Manual
9. V-ACTOR II Specifications and Operation
9.11 CLEANING
CAUTION
Cleaning agents and disinfectants can form an explosive atmosphere. Disconnect the handpiece and the control device
from the mains before starting any cleaning or maintenance work.
ATTENTION
Make absolutely sure that no liquid can seep into the handpiece.
The handpiece must be thoroughly cleaned and disinfected after each therapy session.
• Clean the remaining massage oil off the handpiece using a damp cloth.
• Disinfect the handpiece with a surface disinfectant.
• Clean the shock transmitters thoroughly after each use. Unscrew the shock transmitter screw cap and remove the shock
transmitter insert if necessary.
• The shock transmitters can be cleaned using the usual cleaning agents and disinfectants.
• Clean the shock transmitters each day in an ultrasonic bath.
• Proceed as follows to clean the shock transmitters:
V40 shock transmitter:
• Unscrew the shock transmitter from the handpiece, disassemble the threaded two-part shock transmitter screw cap, remove
the shock transmitter insert from the front cap and take out the front sealing ring. This is cut open so that it can be removed
more readily. The spring element on the shock transmitter insert does not need to be removed.
• Clean all of the parts under running water.
• Clean and disinfect the shock transmitter insert and the sealing ring in an ultrasonic bath.
• For this purpose, use only instrument disinfectants for heat-sensitive, reusable medical devices.
2
1
Fig. 9 - 7 V40 shock transmitter disassembled
1 Spring element
2 Front sealing ring