14
Mobile 2 RPW User Manual
3. Operation
NOTE
The device meets the requirements of the applicable electromagnetic compatibility (EMC) standards EN 60601-1-2.
These requirements are designed to provide reasonable protection against harmful interference in a typical medical
installation. The instrument described here generates and uses high-frequency energy and can emit the same. If not
installed and used in accordance with these instructions, the instrument may cause harmful interference with other devices
in the vicinity. However, there is no guarantee that interference will not occur in a particular installation. If this product
does cause harmful interference with other devices, which can be determined by turning the instrument off and on, the
user is encouraged to try to correct the interference by one or more of the following measures:
•
Reorient or relocate the receiving device.
•
Increase the distance between the devices.
•
Connect the devices to an outlet on a circuit different from that to which the other device is connected.
•
Consult the manufacturer or field service technician for help.
3.2 START-UP
NOTE
Prior to start-up, please refer to chapter 8 and 9 for your handpiece.
•
Switch on the device at the mains switch on the back of the instrument (Fig. 1 - 3/7).
Once the unit has been started, the display automatically shows the last shock frequency setting. The display flashes.
•
To confirm the existing setting, press one of the two arrow keys.
•
To change the existing setting, press one of the two arrow keys. As soon as the display has stopped flashing, the selected shock
frequency can be increased or reduced using the arrow keys.
With the device, the shock wave frequency can be selected in steps.
Operating Mode
Energy/Frequency
R-SW
1.0 - 4.0 bar / max. 21 Hz
R-SW
4.6 bar / max. 15 Hz
R-SW
5.0 bar / max. 12 Hz
V-ACTOR
1.4 - 3.0 bar / max. 31 Hz
V-ACTOR
3.8 bar / max. 22 Hz
V-ACTOR
5.0 bar / max. 13 Hz
•
Set the energy of the shocks to an initial value of 1.5 bar using the dial (Fig. 3 - 1/5). The value is displayed on the pressure display (Fig. 3 - 1/3).