5. Error Messages and Trouble-shooting
5.1 WARNINGS
The following list gives the most important error codes and the actions that you should take if they occur.
Error number
Fault description
Corrective action
Error 1
Memory error
Acknowledge by pressing
the reset button,
continued operation is possible.
Error 10
Trigger button is pressed
during start-up
Release the trigger button, continued
operation is possible.
Error 11
The handpiece is not connected
Connect the handpiece, continued
operation is possible.
Error 12
Internal fault
Acknowledge by pressing
the reset button,
continued operation is possible.
Error 20
Internal fault
Acknowledge by pressing
the reset button,
continued operation is possible.
Error 21
Internal fault
Acknowledge by pressing
the reset button,
continued operation is possible.
Error 22
Problem with update
on USB stick
Acknowledge by pressing the reset button.
Check update on stick,
continued operation is possible.
Error 23
USB stick not inserted
Insert USB stick
Error 24
Problem with update
on USB stick
Acknowledge by pressing the reset button.
Check update on stick,
continued operation is possible.
Error 25
Problem with update
on USB stick
Acknowledge by pressing the reset button.
Check update on stick,
continued operation is possible.
Error 26
No current
software update
Acknowledge by pressing the reset button.
Check update on stick,
continued operation is possible.