Mobile 2 RPW User Manual
8.5.2 OVERHAULING THE HANDPIECE
CAUTION
Danger of injury due to shock triggering when handpiece is open.
●
Disconnect the handpiece from the control unit before replacing the transmitter
ATTENTION
An open-end spanner should be used to release the handpiece shaft when overhauling the handpiece.
• Place the handpiece on a dry, clean and dust-free surface.
• Unscrew the transmitter screw cap from the handpiece.
• Disconnect the handpiece from the control device.
1
2
• Unscrew the shaft from the handpiece and pull it out of the handpiece handle (5).
• Use the supplied open-end spanner for this purpose (5).
4
5
• Remove the tightly fitting guide tube from the shaft. If necessary, use a thin metal rod or
the supplied hexagonal spanner as a pulling tool by inserting it through the openings in
the guide tube.
6
• A corresponding fixture is provided in the handpiece handle to hold the projectile back.
To remove the projectile, hold the handpiece handle with its opening pointing down.
Gently knock the handle against the table surface until the projectile falls out. In the
event that the projectile breaks apart due to overloading, a fragment of the projectile
may be left behind inside the guide tube.
7