• Screw the transmitter screw cap with the required transmitter tight back onto the shaft.
• Make sure that the two cap parts of the D20 and the D35 shock transmitters are scre-
wed firmly in place and that the shock transmitter screw cap is screwed firmly to the
shaft.
15
• Carry out a functional check of the handpiece (see
Chapter 3.3 Functional Checks
).
• Reset the handpiece shock counter on the control device (refer to
Chapter 3.7 Resetting the Handpiece Shock
Counter
).
8.6 MAINTENANCE
Preventive maintenance is not necessarily required.
8.7 DISPOSAL
When disposing of this product, no special measures have to be observed. Please proceed in accordance with applicable country-
specific regulations. After expiration of the service life of the handpiece, please return the instrument to DJO LLC.
8.8 REPAIR
Repair work on defective instruments must only be carried out by personnel suitably authorised by DJO LLC. Use only original parts
designated by STORZ MEDICAL AG and supplied by DJO LLC. The personnel suitably authorised can be from DJO LLC or be repre-
sentatives of DJO LLC agencies and dealers.
8.9 SERVICE LIFE
The R-SW handpiece should be overhauled about every 1 million shocks (see
Chapter 8.5 R-SW Handpiece Overhaul
).
Provided this interval is observed, the average expected service life is approx. 5 million shocks for the handpiece and 1 million shocks
for the transmitters.
Exceeding the service life can be expected to result in a failure of the instruments.
No warranty claims shall be accepted beyond the information given in
Chapter 10 Warranty and Service
.
For information about the service life of your control device, please refer to the
Chapter 4.7 Service Life
.