9. V-ACTOR II Specifications and Operation
9.15 TROUBLE-SHOOTING
Fault description
Possible cause
Corrective action
No compressed air sup-
ply
Leaks in handpiece ca-
ble, or cable not prop-
erly connected
Check the cable and tube
connections and replace
them, if necessary.
No power output
Handpiece defective
Replace the handpiece.
CAUTION
Disconnect the handpiece from the control device before starting any maintenance work!