9. V-ACTOR II Specifications and Operation
V25 shock transmitter:
• Unscrew the shock transmitter from the handpiece and remove the shock transmitter insert from the front cap. The spring
element on the shock transmitter insert does not need to be removed.
• Clean all of the parts under running water.
• Clean and disinfect the shock transmitter insert and the sealing ring in an ultrasonic bath.
• For this purpose, use only instrument disinfectants for heat-sensitive, reusable medical devices.
1
Fig. 9 - 8 V25 shock transmitter
1 Spring element
Fig. 9 - 9 Fig Ultrasonic bath with V-ACTOR shock transmitters
9.12 DISPOSAL
No special precautions need be taken when disposing of the V-ACTOR ll handpiece. Please proceed in accordance
with applicable country-specific regulations. The instrument must be disposed of as waste electronic equipment at
the end of its service life.
9.13 REPAIR
Repair work on defective instruments must only be carried out by personnel suitably authorised by DJO LLC. Use only original parts
designated by STORZ MEDICAL AG and supplied by DJO LLC. The personnel suitably authorised can be from DJO LLC or be repre-
sentatives of DJO LLC agencies and dealers.
9.14 SERVICE LIFE
The average expected service life of the handpiece is approx. 5 million pulses.
Exceeding the service life can be expected to result in a failure of the instruments.
No warranty claims shall be accepted beyond the information given in
Chapter 10 Warranty
.
For information about the service life of your control device, please refer to
Chapter 4.7 Service Life
.