Intelect® Advanced sEMG and sEMG + Stim Module
49
TROUBLESHOOTING
ERROR CODES (continued)
Code
Number
Type
Message
Probable Cause
Possible Remedies
145
Warning
Patient Data Card button on Home screen was pressed with no
Patient Data Card installed into system port and no treatment
currently being performed.
Properly insert a Patient Data Card, set up and perform the treatment, and save data to Patient Data
Card.
In the event that an Error message or Warning appears beginning with
a 2 or 3, immediately stop all use of the system and contact the dealer
or DJO for service. Errors and Warnings in these categories indicate an
internal problem with the system that must be tested by DJO or a Field
Service Technician certified by DJO before any further operation or use
of the system.
Use of a system that indicates an Error or Warning in these categories
may pose a risk of injury to the patient, user, or may cause extensive
internal damage to the system.
WARNING