Intelect® Advanced sEMG and sEMG + Stim Module
47
TROUBLESHOOTING
ERROR CODES (continued)
Code
Number
Type
Message
Probable Cause
Possible Remedies
130
Warning
Another treatment is running while attempting to set up and
perform a Laser Therapy treatment.
A.
Allow existing treatment to complete before starting Laser Therapy.
B.
If no other treatment is running, reset Therapy System by turning main power switch
Off and On.
131
Warning
Treatment Room Door Lockout is breached.
A.
Make certain Treatment Room Door is completely closed.
B.
Make certain the Lockout cable is connected to the system.
C.
Replace Lockout to System cable with a known good cable.
D.
Contact department responsible for installation of the Treatment Room Door Lockout mechanism for
maintenance or repair.
E.
If problem continues, send Laser Module to factory for service.
132
Warning
Attempted to start a laser treatment but no Laser Applicator is
plugged in.
A.
Connect desired Laser Applicator to the system.
B.
If Applicator is connected, reset Therapy System by turning main power switch Off and On.
C.
Connect a known good Laser Applicator.
D.
If problem continues, send Laser Module to factory for service.
133
Warning
Laser Applicator became unplugged while performing a laser
treatment.
A.
Connect desired Laser Applicator to the system.
B.
If Laser Applicator is connected, reset Therapy System by turning main power switch
Off and On.
C.
Connect a known good Laser Applicator.
D.
If problem continues, send Laser Module to factory for service.
134
Warning
Entered incorrect laser PIN.
A.
Enter correct Laser PIN number.
B.
If problem continues, send Laser Module to factory for service.
135
Warning
Control Board software upgrade warning.
Upgrade Control Board software to latest version. Contact dealer or DJO for latest software upgrade and
instructions.
136
Warning
Stim Board Main software upgrade warning.
Upgrade Stim Board software to latest version. Contact dealer or DJO for latest software upgrade and
instructions.