Intelect® Advanced sEMG and sEMG + Stim Module
14
GENERAL INFORMATION
INSTALLATION & REMOVAL
The Intelect Advanced (Color Series) Electrotherapy and
Combination Therapy Systems have, as standard equipment, an
sEMG Module factory installed.
A second sEMG Module may be installed onto the Optional
Channel 3/4 Electrotherapy Module only. The sEMG Module will
not install to any other optional modules such as the Battery or
Laser Modules.
The sEMG Module is available as an option for the Intelect
Advanced (Monochromatic) Therapy System.
These instructions assume that there is no sEMG Module installed
on the Therapy System in order to provide the proper installation
instructions for future use should it become necessary.
sEMG Module
If installing the sEMG Module on a Therapy System equipped
with a Battery, Laser, or the Channel 3/4 Electrotherapy Module
and it is desired to install the sEMG Module onto the Therapy
System, it will be necessary to remove the Battery, Laser, or
Channel 3/4 Electrotherapy Module. Refer to the respective
Module User Manual for Instructions.
Installing the sEMG Module on the System will make the sEMG
and sEMG + Stim modalities available for Channels 1/2. If installed
on the Optional Channel 3/4 Electrotherapy Module, sEMG and
sEMG + Stim modalities will be available on Channels 3/4.
If two sEMG Modules are installed, one on the Therapy System
and one on the Channel 3/4 Electrotherapy Module, sEMG and
sEMG + Stim modalities will be available to all four Channels.
However, only two channels at a time may be used; either
Channels 1/2 or Channels 3/4.
This section provides general instructions to prepare the Therapy
System and the optional Channel 3/4 Electrotherapy Module for
installation of the sEMG and sEMG + Stim Module(s).
Therapy System
WARNING
Disconnect the system from the power source (outlet or remove
battery module if installed) before attempting any maintenance,
installation, removal, or replacement procedures to prevent electrical
shock and possible damage to system.