Intelect® Advanced sEMG and sEMG + Stim Module
15
THERAPY SYSTEM PREPARATION
INSTALLATION & REMOVAL
Disconnect System Mains
Disconnect the Power Cord from the
power supply.
Remove the Rear Panel and disconnect the
Power Cord from the System.
Remove the Front Access Panel and discon-
nect all accessories.
REAR
ACCESS
PANEL
DISCONNECT
POWER CORD
Disconnect Leads and Accessories
To remove Therapy System from Cart, refer
to the Therapy System Cart User Manual.
DISCONNECT ALL ACCESSORIES
Remove Therapy System from Cart
Place the System upside down on a
smooth surface. Place a soft, clean fabric or
cushion under the System Display Lens to
prevent damage.
Remove Module
Remove Channel 3/4 Electrotherapy,
Battery, or Laser Module if equipped. Refer
to the respective User Manual for the
Module being removed.
WARNING
Disconnect the system from the power source (outlet or remove
battery module if installed) before attempting any maintenance,
installation, removal, or replacement procedures to prevent electrical
shock and possible damage to system.