Intelect® Advanced sEMG and sEMG + Stim Module
13
The High Voltage Pulsed Current (HVPC) has a very brief pulse
duration characterized by 2 distinct peaks delivered at high
voltage. The waveform is monophasic (current flows in one
direction only). The high voltage causes a decreased skin
resistance making the current comfortable and easy to tolerate.
Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrodes or Probe
Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0-500 V
Polarity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Positive or Negative
Ramp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.5 sec, 1 sec, 2 sec, 5 sec
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Peak Current or Volts
Sweep . . . . . . . . . . . . . .Continuous, 80/120 pps, 1/120 pps, 1/10 pps
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-120 pps
Cycle Time . . . . . .5/5, 4/12, 10/10, 10/20, 10/30, 10/50, Continuous
Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 Minutes
Available on Channels . . . . . . . . . . . . . . . . . . . . . . . . 1 & 2, 3 & 4 Option
*CC= Constant Current
CV= Constant Voltage
WAVEFORM SPECIFICATIONS (continued)
SPECIFICATIONS
High Voltage Pulsed Current (HVPC)
VMS is a symmetrical biphasic waveform with a 100 µsec interphase
interval. Because the pulse is relatively short, the waveform has a low
skin load, making it suitable for applications requiring high intensities,
such as in muscle strengthening protocols.
Output Mode: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Electrodes
Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-255 mA
Channel Mode . . . . . . . . . . . . . . . . . . . . . . . . . Single, Reciprocal, Co-Contract
Phase Duration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20-1000µsec
Mode Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CC or CV*
Anti-Fatigue . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Off or On
Set Intensity . . . . . . . . . . . . . . . . . . . . Individual Channel Intensity Setting in
Reciprocal and Co-Contract modes
Cycle Time . . . . . . . . . . .Continuous, 5/5, 4/12, 10/10, 10/20, 10/30, 10/50
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-200 pps
Ramp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0.5 sec, 1 sec, 2 sec, 5 sec
Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 minutes
Available on Channels . . . . . . . . . . . . . . . . . . . . . . . . 1 & 2, 3 & 4 Option
VMS
TM