Intelect® Advanced sEMG and sEMG + Stim Module
29
OPERATION
S
EMG THERAPY SET UP (continued)
Set Audio Type
Audio sets the type of sound heard when
the muscle contraction reaches the Alarm
setting.
Press the Audio button until the desired
audio type is displayed in the Audio icon.
The available selections are;
Constant-
Continuous sound. ________
Pulsed-
Fast, short beeps. . . . . . . . . . . . . . .
Dynamic-
Slow, long beeps. _ _ _ _ _ _
ALARM
BUTTON
NOTE:
If volume is Off, no Audio type will be
heard.
Three options are available for setting the
sEMG modality target:
Max-
Maximum of muscle contractions in
10 seconds.
Avg-
Average of maximums achieved in
15 seconds of muscle contractions.
Manual-
Set Target manually.
Press the Target button until the prescribed
target option is displayed in the Target icon.
Select Target Option
TARGET
BUTTON
Setting Max Target
Make certain Target Max is displayed in the
Target icon.
Press the Capture Target button. Have the
patient begin contracting the muscle and
press the Begin Capture button. Patient
should contract the muscle as hard as
possible during the preset 10 second
contraction period.
CAPTURE TARGET
BUTTON
BEGIN CAPTURE
BUTTON
NOTE:
The capture may be stopped by pressing
the End Capture Button. The System will
then select the maximum contraction level
achieved during the preset 10 second
contraction period.