Intelect® Advanced sEMG and sEMG + Stim Module
39
OPERATION
S
EMG + STIM THERAPY SET UP (continued)
Start sEMG Therapy Session
Have patient perform contractions as
prescribed.
When the contraction reaches the target,
the alarm will sound, the screen will
change, and Stim will begin. The patient
should hold the contraction throughout
the short period of Stim.
After all parameters for sEMG and Stim have
been completed, press the Start sEMG +
Stim button.
START SEMG+STIM
BUTTON
NOTE:
The Stim Intensity may be adjusted only
during the Stim portion of the sEMG
+ Stim therapy session.
Set Auto Feature
The sEMG + Stim Auto feature is designed
to work in conjunction with the Target
settings. When On, the Auto feature will
automatically adjust the maximum target
as the patient fatigues.
Press the Auto button until Off or On is
displayed in the Auto icon.
AUTO
BUTTON