Intelect® Advanced sEMG and sEMG + Stim Module
48
TROUBLESHOOTING
ERROR CODES (continued)
Code
Number
Type
Message
Probable Cause
Possible Remedies
137
Warning
Stim Board Main software upgrade warning.
Upgrade Stim Board sSoftware to latest version. Contact dealer or DJO for latest software upgrade and
instructions.
138
Warning
Ultrasound Board software upgrade warning.
Upgrade Ultrasound Board software to latest version. Contact dealer or DJO for latest software upgrade
and instructions.
139
Warning
Laser Board software upgrade warning.
Upgrade Laser Board software to latest version. Contact dealer or DJO for latest software upgrade and
instructions.
140
Warning
MMC software upgrade warning.
Upgrade MMC software to latest version. Contact dealer or DJO for latest software upgrade and
instructions.
141
Warning
Battery Module software upgrade warning.
Upgrade Battery software to latest version. Contact dealer or DJO for latest software upgrade and
instructions.
142
Warning
A Laser Protocol was selected but no Laser Module is installed
on system.
Install Laser Module to Therapy System. Refer to Laser User Manual for installation instructions.
143
Warning
A Laser Protocol was selected but no Laser Applicator connected
to system.
A.
Connect proper Laser Applicator to the system.
B.
If Laser Applicator is connected, reset Therapy System by turning main power switch
Off and On.
C.
Connect a known good Laser Applicator.
D.
If problem continues, send Laser Module to factory for service.
144
Warning
Wrong Laser Applicator connected to system for the protocol
selected.
A.
Connect correct Laser Applicator to the system.
B.
If Applicator is connected, reset Therapy System by turning main power switch Off and On.
C.
Connect a known good Laser Applicator.
D.
If problem continues, send Laser Module to factory for service.