Intelect® Advanced sEMG and sEMG + Stim Module
18
INSTALLING
S
EMG MODULE (continued)
INSTALLATION & REMOVAL
Install Cables and Accessories
Install all cables and accessories to the Front
Access Panel. Refer to
page 9
for Symbol
Definitions.
Plug Power Cord into an
approved outlet.
Turn the System On
using the On/Off
Switch. The System will
automatically recognize
the added Module and
display a configuration
change message.
Read and carefully
follow the instructions
on the Screen.
CONNECT ALL CABLES
AND ACCESSORIES
Apply Mains Power
Verify that the Module installed is the Module displayed in the message
BEFORE pressing the START Button. If it is not, DO NOT press the START
Button. Turn the System OFF and back ON. If the problem continues, call
the selling dealer or DJO Technical Support immediately.
DO NOT USE THE SYSTEM until all necessary repairs are made by a
Technician certified by DJO. If use is attempted before repairs are made,
the System may operate unpredictably and has the potential of causing
injury to the patient or damage to the System's internal components.
WARNING