Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
29
6 REMOVAL & REPLACEMENT
A. Part Number
.......................................................... 27040
B. Tools and Equipment Required
#1 Phillips Screwdriver
C. Probe Spring Holder Removal and
Replacement
1. Remove the Sound Head per
Section 6.8
part
C.
2. Remove the brass Probe Spring and inspect
for damage. Retain the Probe Spring for
installation if undamaged.
Refer to
Figure
6.25.
3. Using a # 1 Phillips screwdriver remove the
screw securing the Probe Spring Holder to
the Applicator housing.
Refer to
Figure
6.25.
4.
Reverse
steps 1 through 3
for installation of
the replacement spring.
5. Recalibrate the Applicator.
NOTE:
Do not over tighten the screw holding the
Probe Spring Holder. A cracked or damaged
Probe Spring Holder will affect the operation
of the Applicator.
•
6.9 APPLICATOR PROBE SPRING HOLDER
Applicator must be calibrated after removal
and replacement of any part, using an Ohmic
UPM DT 10 or UPM DT 100.
Failure to calibrate a repaired Applicator, using
the proper equipment, may result in serious
injury to the patient.
FIGURE 6.25
BRASS
PROBE SPRING
SCREW
PROBE SPRING
HOLDER