![Chattanooga Group Intelect TRANSPORT Vectra Genisys 2759 Скачать руководство пользователя страница 28](http://html.mh-extra.com/html/chattanooga-group/intelect-transport-vectra-genisys-2759/intelect-transport-vectra-genisys-2759_service-manual_2592343028.webp)
Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
26
6.5 LCD DISPLAY
A. Part Number
.......................................................... 27264
B. Tools and Equipment Required
#1 Phillips Screwdriver
Orange stick
C. Ultrasound Board
1. Separate top from bottom.
Refer to 6.1, part C.
2. Remove Ultrasound Board.
Refer to 6.4, part C.
3. Using a # 1 Phillips screwdriver remove the 4
LCD retaining screws.
Refer to Figure 6.16.
4. With an orange stick or flat edged piece of
plastic gently push in and up on each end of
the LCD retaining brackets to free the locking
tabs from the Control Board.
Refer to Figure 6.16.
5. Set retaining brackets aside for reassembly.
6. Carefully remove the LCD from the Control
Board being careful not to damage the LCD
header
pins.
Refer to Figure 6.17 .
7. Install the LCD by reversing
steps 1 through 6
.
NOTE:
Make certain to align the LCD header pins with
the top row on the Control Board LCD
connector.
Refer to Figure 6.17.
•
•
FIGURE 6.16
Unplug the unit from the power source and/or
remove battery before attempting any removal
or replacement procedures to prevent electrical
shock.
FIGURE 6.17
2 SCREWS
EACH END
RELEASE RETAINING
BRACKET TABS
REMOVE LCD FROM
CONTROL BOARD
6 REMOVAL & REPLACEMENT