Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
28
6 REMOVAL & REPLACEMENT
FIGURE 6.21
FIGURE 6.22
A. Part Numbers (Applicators)
Intelect
Applicators
1 cm² .............................................................................. 27381
2
cm²
.............................................................................. 27382
5
cm²
.............................................................................. 27383
10
cm²
........................................................................... 27384
Genisys Applicators
1 cm² .............................................................................. 27333
2
cm²
.............................................................................. 27334
5
cm²
.............................................................................. 27335
10
cm²
........................................................................... 27336
Sound
Heads
1 cm² .............................................................................. 27126
2
cm²
.............................................................................. 27127
5
cm²
.............................................................................. 27128
10
cm²
........................................................................... 27129
B. Tools and Equipment Required
None
C. Sound Head Removal and Replacement
NOTE:
The Transducer is contained within the
Sound Head of the Applicator.
1. Unscrew the Sound Head from the body of
the
Applicator.
Refer to Figure 6.21.
2. After removing the Sound Head inspect
the plastic Probe Spring Holder in the top of
the Applicator Assembly for cracks or
discoloration.
Refer to Figure 6.22.
NOTE:
If the Probe Spring Holder is dark or contains
crack it should be replaced.
Refer to Section
6.9, part C.
3. Reverse steps 1 and 2 to install a
replacement Sound Head.
4. Recalibrate the Applicator following the
calibration
procedure
in
Section 7, part B.
•
6.8 APPLICATOR SOUND HEAD
Applicator must be calibrated after removal
and replacement of any part, using an Ohmic
UPM DT 10 or UPM DT 100.
Failure to calibrate a repaired Applicator, using
the proper equipment, may result in serious
injury to the patient.
PROBE
SPRING
HOLDER