Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
17
5 TROUBLESHOOTING
5.9 ULTRASOUND TESTS
A. Equipment Required for 5.9 and 5.10
Degassed Water.
Refer to page 12
for
degassed water recipes
Ohmic Instruments UPM DT 10 or UPM DT 100
Ultrasound Power Meter
Known good Ultrasound Applicator
NOTE
:
Use any Ultrasound Applicator supplied with a
Chattanooga Group unit for this test.
B. Ultrasound Applicator Identification
Test
Procedures
1. Without Ultrasound Applicator installed,
turn unit on.
2. Look at the Display center. The Applicator
not detected icon should appear.
Refer to
Figure
5.4
.
3. Connect the Applicator into Applicator
Receptacle.
Refer to Figure 5.5.
Watch
Applicator LED while connecting to system.
The LED should flash green five times.
NOTE:
The flashing green light indicates that the
Applicator is programmed and calibrated.
4. Look at the display center. The Applicator
detected icon should appear.
Refer to
Figure
5.6
.
C. Ultrasound Applicator Identification
Test
Results
1. Unit operates as described in steps 2 and 4.
a) Unit passed test.
2. “NO CAL” icon displays center of system
Display.
a) Applicator not calibrated.
b) Possible bad Applicator Cable. Re-test
with known good Applicator.
c) Possible bad Ultrasound Board.
3. Not detected icon displayed after ten
seconds of Applicator being plugged in.
a) Possible bad Applicator or Applicator
Cable. Re-test with known good
Applicator.
b) Possible bad Ultrasound Board.
•
•
•
FIGURE 5.4
FIGURE 5.5
FIGURE 5.6