13
5.3 ELECTRICAL SAFETY
The Intelect Transport/Vectra Genisys Ultrasound
Therapy Systems has been tested to IEC/EN
60601-1, Standard for Safety for Medical
Equipment.
NOTE:
NOTE:
This device complies with current leakage,
This device complies with current leakage,
ground continuity, and dielectric withstand
ground continuity, and dielectric withstand
(Hi-Pot) limits as prescribed by IEC/EN/UL
(Hi-Pot) limits as prescribed by IEC/EN/UL
60601-1 and CSA/CAN 601.1 Medical Electrical,
60601-1 and CSA/CAN 601.1 Medical Electrical,
Part 1: General Requirements for Safety.
Part 1: General Requirements for Safety.
Facility, local and national limits and test
Facility, local and national limits and test
methods may vary.
methods may vary.
Power
Requirements
Models: 2782 and 2759 . . . . . . Input : 100-240VAC
(Serial numbers 1000 and above) 75 VA, 50/60 Hz
5.4 LEAKAGE
Conduct all necessary leakage tests as required
per NFPA 99 (National Fire Protection Association)
“Health Care Facility” standards.
5 TROUBLESHOOTING
Unit failing Dielectric Withstand or Leakage tests
could indicate serious internal problems.
Do not place unit back into service! Send unit to
factory for repair. Do not attempt to repair.
9700675
Listed by Intertek Testing Services NA Inc.
Conforms to UL Standard 60601-1
Certified to CAN/CSA Standard C22.2 No.
601.1-M90 w/A2
Compliant to IEC/EN 60601-1,
IEC 60601-1-2, 60601-2-5