Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
24
6.3 POWER SUPPLY (CONTINUED)
6. Remove Power Supply from system.
7. Carefully remove the copper tape securing the
shield to the Power Supply. Remove from the
inside of the Power Supply and retain the
Shield for installation of the replacement.
NOTE:
The black side of the shield is non conductive
and should always face the Power Supply.
8.
Reverse
steps 4 - 7
to install the replacement.
Refer to Figure 6.8
to correctly route Power
Supply Harness. Make certain the Power Supply
sits onto alignment pins as shown in
Figure 6.11.
FIGURE 6.11
Handle Power Supply Shield with caution.
Penetration of the Shield may result in
damage to the unit and/or personal injury.
6 REMOVAL & REPLACEMENT
POWER SUPPLY
ALIGNMENT PINS