Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
27
6.6 CONTROL BOARD
A. Part Number
.......................................................... 27268
B. Tools and Equipment Required
#1 Phillips Screwdriver
Orange Stick
C. Control Board
1. Separate top from bottom.
Refer to 6.1, part C
.
2. Remove Ultrasound Board.
Refer to 6.4, part C.
3. Remove LCD from Control Board.
Refer to 6.5,
part
C.
4. Remove the 7 retaining screws securing the
Control Board to the housing.
Refer to
Figure
6.18.
5. Using an orange stick or flat edged piece of
plastic, carefully push locking tabs while lifting
the Control Board to release Control Board from
housing.
Refer to Figure 6.19.
6. Remove the LCD Contrast Knob and retain for
installation on the new Control Board.
Refer to
Figure
6.20.
7. Install the replacement Control Board by
reversing
steps 1 - 6.
NOTE:
Do not overtighten the screws. Overtightening
will damage the threads of the brass inserts.
•
•
FIGURE 6.18
Unplug the unit from the power source and/or
remove battery before attempting any removal
or replacement procedures to prevent electrical
shock.
FIGURE 6.19
FIGURE 6.20
6 REMOVAL & REPLACEMENT
6.7 KEYMATS
A. Part Number
.......................................................... 27368
B. Tools and Equipment Required
#1 Phillips Screwdriver
Flat Blade Screwdriver
C. Keymat
NOTE:
Follow
6.6 steps 1 through 5
, for Control Board
removal. When the Control Board is removed lift
the Keymats out and replace.
Keymats set includes the On/Off button.
•
•
RETAINING SCREWS
PUSH TABS IN AND
PULL UP ON BOARD
SIMULTANEOUSLY
CONTRAST KNOB