Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
30
6 REMOVAL & REPLACEMENT
A. Part Numbers
1
cm²
.............................................................................. 27600
2
cm²
.............................................................................. 27071
5
cm²
.............................................................................. 27073
10
cm²
........................................................................... 27075
B. Tools and Equipment Required
#1 Phillips Screwdriver
C. PC Board Removal and Replacement
1. Remove the Sound Head per
Section 6.8,
part
C.
2. Remove the Probe Spring and Probe Spring
Holder
per
Section 6.9, part C
.
3. Remove the Applicator Cable per
Section
6.11, part C.
4. Pull the PC Board from the Applicator
housing.
Refer to Figure 6.24.
NOTE:
If necessary needle nose pliers can be used
to grasp the edge of the PC Board for
removal.
6. Reverse steps 1 through 4 for installation of
the replacement PC Board.
7. Recalibrate the Applicator.
NOTE:
When replacing the PC Board it should be
oriented with the Cable Connector toward
the inside of the Applicator Handle. Push the
PC Board as far into the Applicator Handle as
possible, with the Cable end fitting into the
two PC Board slots.
Refer to Figure 6.25.
The Sound Head end of the PC Board fits
against the screw standoff for the Probe
Spring
Holder.
Refer to Figure 6.26.
•
6.10 APPLICATOR PC BOARD
Applicator must be calibrated after removal
and replacement of any part, using an Ohmic
UPM DT 10 or UPM DT 100.
Failure to calibrate a repaired Applicator, using
the proper equipment, may result in serious
injury to the patient.
FIGURE 6.24
FIGURE 6.25
GROOVES FOR PC BOARD
FIGURE 6.26
PROBE SPRING
HOLDER STANDOFF