Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
9
A.
Ultrasound
Frequency . . . . . . . . 1 MHz, ± 5%; 3.3 MHz, ± 5%
Duty Cycles . . . 10%, 20%, 50%, and Continuous
Pulse Frequency . . . . . . . . . . . . . . . . . . . . . . . . .100 Hz
Pulse Duration . . 1 mSec, ± 20%; 2 mSec, ± 20%
5 mSec, ± 20%
Output
Power
10
cm
2
Crystal . . . . . . . . . . . . 0-20 Watts at 1 MHz,
0-10 Watts at 3.3 MHz
5
cm
2
Crystal . . . . . . . . 0-10 Watts, 1 and 3.3 MHz
2
cm
2
Crystal . . . . . . . . . .0-4 Watts, 1 and 3.3 MHz
1
cm
2
Crystal . . . . . . . . . . .0-2 Watts, 3.3 MHz Only
Amplitude . 0 - 2.5 w/cm
2
in Continuous mode,
0-3 w/cm
2
in Duty Cycle modes
Output accuracy . . . . . . . . . .± 20% above 10% of
maximum
Temporal Peak to Average Ratios:
2:1, ± 20%, at 50% Duty Cycle
5:1, ± 20%, at 20% Duty Cycle
9:1, ± 20%, at 10% Duty Cycle
Beam Non uniformity Ratio . . . .. . . . . . . . . . 5.0 : 1
maximum
Beam Type . . . . . . . . . . . . . . . . . . . . . . . . . Collimating
Effective Radiating Areas
10
cm
2
Crystal - 6.8 cm
2
- 10.0 cm
2
5
cm
2
Crystal - 3.5 cm
2
- 5.0 cm
2
2
cm
2
Crystal - 1.4 cm
2
- 2.0 cm
2
1
cm
2
Crystal - 0.7 cm
2
- 1.0 cm
2
Treatment Time . . . . . . . . . . . . . . . . . . 1-30 Minutes
Do not apply the Ultrasound Applicator to
the patient during the Head Warming period.
Applicator must remain in applicator hook during
the Head Warming period.
4.2 INTELECT TRANSPORT/VECTRA GENISYS ULTRASOUND THERAPY
SYSTEMS
ULTRASOUND SPECIFICATIONS
This section provides the necessary ultrasound specifications to aid in troubleshooting.
Refer to these specifications as necessary when troubleshooting the ultrasound PC Board and Applicators.
B. Head Warming Feature Specifications
The Head Warming feature of an Intelect
Transport/Vectra Genisys Ultrasound Therapy
System utilizes Ultrasound output resulting in
warming of the Applicator to increase patient
comfort.
With Head Warming enabled, ultrasound is
emitted without pressing the Start button. The
Applicator LED will not illuminate during the
Head Warming period. Display will indicate
"Warming".
Output . . . .
0 - 50% Cycling of maximum power
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . .3.3 MHz
Sound Head Temperature . . . . . . 85 °F - 110 °F
(29.4 °C - 43.3 °C)
4 SPECIFICATIONS