Intelect TranSport®/Vectra Genisys® Ultrasound Therapy Systems
11
ERROR
CODE
ERROR
TYPE
DEFINITION
PROBABLE CAUSES
POSSIBLE REMEDY
USER CORRECTABLE WARNING MESSAGES
300
301
302
303
304
305
CRIT_ERR_US_BOARD
CRIT_ERR_US_BOARD
NOT_DETECTED
CRIT_ERR_US_RESET
CRIT_ERR_US_BOARD_
READ
CRIT_ERR_US_BOARD_
WRITE
CRIT_ERR_US_BOARD_
CALIBRATION_ERROR
Some type of critical Ultrasound Board
error has occurred
No Ultrasound Board is detected in unit
Ultrasound Board has reset for some
unknown reason
Error reading from Ultrasound Board
Error writing to Ultrasound Board
Error calibrating Applicator
Ultrasound Board to Control Board Connecting
Header loose or disconnected
Bad Ultrasound Board
Bad Control Board
1. Reseat Connecting Header into Control Board
2. Replace Ultrasound Board
3
.
Replace Control Board
NONE
No indication of problem and no
ultrasound output from the Applicator
Probe Spring Holder cracked or broken
Replace Probe Spring Holder
NONE
Interrupted ultrasound output from
Applicator
Bad Applicator Cable
Replace Applicator Cable
5.1 INTELECT TRANSPORT/VECTRA GENISYS ULTRASOUND THERAPY
SYSTEMS
ERROR
MESSAGES CONTINUED
5 TROUBLESHOOTING