Page 111
AGND
AGND
SIGNAL
SIGNAL
C1
4700uF
R1
3k9
SIGNAL
GENERATOR
INPUT SOCKET
TRANSDUCER
Specification
Whole Body
(KD1007)
Whole Body
(KD1009)
Output Voltage
100mV/g
±5%
1000mV/g
±5%
Operating Range
±50g
±3.6g
Frequency Response
0.5 to 3000 Hz
±5%
0.1 to 125 Hz
±5%
Resonant Frequency
25000Hz
5000Hz
Weight
227 Grams
550 Grams
Operating Temperature
Range
-50°C to 70°C
-60°F to 160°F
-10°C to 50°C
14°F to 122°F
Thermal Sensitivity
Coefficient
0.06% / °C
0.03% / °F
0.012dB / °C
Linearity
±1%
±1%
Electrical Noise Floor
0.0007g
< 316
μ
V rms
Transverse Sensitivity
5%
5%
Maximum Shock
1500g
100g
Seal Hermetic
Hermetic
Electrical Signal Input
Electrical signals at frequencies >1Hz can be applied to the Pro-DX Excieo range
of Vibration Meters by interfacing a suitable signal generator with an output
impedance of 600
Ω
to the 5 pin Input Lemo Socket, type (EGG.1K.304.CLL).
Each individual axis shall be subject to the following circuitry to create the
required d.c offset voltage and current (for type ‘S’ mode Excieo meters connect
only to the ‘X’ Axis). See
Transducer Input
in the
Technical Specification,
Chapter 10
for socket wiring information.
Maximum Electrical Signal Input For No Damage
18 Volts (Peak to Peak)