Page 21
Chapter 5
Real Time Operation
Real Time operation is extremely useful for pre-check range Overload / Under
range condiditons and can be identified by having a Duration value that does not
flash and is set to the value 00:00:00.
Display Modes
Depending on the variation of meter purchased, the following display modes are
available: -
HARM (H)
SINGLE Axis (S)
Multi Function (M)
HARM
S
SINGLE
HARM
WBV
SINGLE
CUSTOM
To switch between the modes of operation for multi function meters press
Soft
Key 1
labelled
MODE
and a popup menu will appear with a list of selectable
display modes. Select which display mode you require by pressing the
appropriate number on your keypad or press
Soft Key 1
labelled
CANCEL
to
cancel the operation and return to the previous screen. Dedicated
HARM
and
SINGLE
axis meters do not have this facility.
It is essential that whichever display mode is used for measuring vibration, that
the correct transducer is attached to your meter. Mismatched display modes
and transducers, or accelerometer sensitivities will give incorrect
measurements.
09:00:30
26:08:11
Aeq
HARM
Range: 0.10 – 500 ms
-2
>
Duration: 00:00:00
< > to change parameter
Y-Wh:
0.124
X-Wh:
0.123
Z-Wh:
0.125
RESET SETUP
MODE RANGE
Figure 12 – Display Mode Selection
09:00:32
26:08:11
Aeq
HARM
Range: 0.10 – 500 ms
-2
>
Duration: 00:00:00
< > to change parameter
Y-Wh:
0.124
X-Wh:
0.123
Z-Wh:
0.125
RESET SETUP
MODE RANGE
1. HARM
2. WBV
3. Single
4. Custom