Page 78
Press either
1
,
2
or
3
to enter the axis coefficients for the X, Y or Z axis
respectively or just
1
for SINGLE axis mode. The axis coefficients are multiplying
factors applied to the instrument to compensate the output of the specific
transducer being used. The axis coefficients can be found on the instrument
calibration certificate.
Enter the coefficient figure for the chosen axis using the numeric keypad. Use
the
Left Arrow
key
to delete unwanted input or press the
OK
key to accept
the inputted value. Repeat for all axes and modes as required. The valid
coefficient range is 0.8000 to 2.0000, any other inputted value will return an
‘
Out of range
’ error message.
Undertaking automatic calibrations will overwrite the existing coefficients with
new ones calculated from the automatic calibration. To return to factory
settings then use the coefficients supplied on your calibration certificate
Once completed or at any stage press
Soft Key 1
labelled
EXIT
to return to the
Calibration Selection Screen.