Page 106
Measured electrically on the 1000 m/s² range of the X-axis with a reference
frequency of 80 Hz at an approximate level of 900 m/s with Fl (LIN) weighting
applied. All errors and measurements are in dB.
Frequency
(Hz)
Wh
Fc
Tolerance
Ws
Tolerance
1 -0.3
0.0
+2,
-
∞
2.0 +4,
-
∞
2 0.1
0.1
+2,
-
∞
0.3 +4,
-
∞
4 0.1
0.2
+2,
-
∞
0.1 +4,
-
∞
8
0.1
0.1
±2
-1.0
+1 / -2
10 0.1
0.1 ±1
-0.6 ±1
20 0.0
0.0 ±1
0.3 ±1
40 0.0
0.1 ±1
0.1 ±1
80 -0.1
0.0 ±1
0.0 ±1
100 0.0
0.0 ±1 0.0 ±1
200 0.1
0.1 ±1 0.1 ±1
400 0.0
0.1 ±2 0.3 ±1
1000
0.0
0.1
±2
0.2
+1 / -2
2000 -1.2 -0.7 +2,
-
∞
1.3 ±4
4000 -6.0 -4.0 +2,
-
∞
-1.2 ±4
Measured electrically on the 1000 m/s² range of the X-axis with a reference
frequency of 80 Hz on Fl (LIN) for Fl, Fb filters and 8 Hz on Fl (LIN) for Fa, Fk
filters, all with an approximate level of 900 m/s. All errors and measurements
are in dB. No tolerance requirements exist for Fl or Fk.
Frequency
(Hz)
Fb
Tolerance
Fa
Tolerance
Fl
Fk
0.1 -3.4 +2,
-
∞
-3.5
+2,
-
∞
-2.9
-3.0
0.2 -0.9 +2,
-
∞
-1.0
+2,
-
∞
-0.9
-0.9
0.4 -0.2 ±1
-0.3 ±1
-0.2
-0.3
1.0 0.0 ±1
-0.1 ±1
0.0
-0.1
2.0 0.1 ±1
0.0 ±1
0.1
0.0
4.0 0.1 ±1
0.0 ±1
0.1
-0.1
8.0 0.1 ±1
0.0 ±1
0.1
-1.4
10 0.1 ±1
0.0 ±1
0.1
-2.9
20 0.0 ±1
0.0 ±1
0.0
-17.6
40 0.1 ±1
0.0 ±1
0.1
-38.8
80 0.0 ±1
-0.1 ±1
REF
-43.6
100 0.0 ±1
-0.1 ±1
0.0
-43.6
200 0.1 ±1
0.0 +2,
-
∞
0.0
-43.6
400 0.1 ±1
-0.1 +2,
-
∞
0.1
-43.6
1000 0.1 ±2
0.1
2000 -0.7 ±2
0.2
4000 -4.0 +2,
-
∞
0.2