Page 77
Manual Calibration
Manual Calibration is selected when you wish to manually enter individual axis
coefficients using the numeric keypad on your instrument. These figures are
located on your instruments calibration certificate. No vibration calibrator is
required for Manual Calibration.
Choosing option 2 from the Calibration Selection Screen will display the
MANUAL CAL
screen shown below. The displayed mode may vary depending
on the instrument version and/or previous selections and also for GA2005H
and GA2005S Excieo models, MODE selection using
Soft Key 4
is not possible.
For the multi-purpose Excieo variation use
Soft Key 4
labelled
MODE
to display a
popup which allows the selection of the available modes. Press
Soft Key 4
labelled
CANCEL
to return to the previous screen without changing the
calibration mode otherwise press
1
,
2
,
3
or
4
depending on which mode /
transducer you wish to manually calibrate.
Figure 89 – Manual Calibration (Type H)
Figure 90 – Manual Calibration (Type M)
Figure 91 – Manual Calibration Mode Selection